![Agilent Technologies Agilent 86120B User Manual Download Page 60](http://html.mh-extra.com/html/agilent-technologies/agilent-86120b/agilent-86120b_user-manual_2867999060.webp)
2-22
Using the Multi-Wavelength Meter
Measuring Laser Separation
To measure channel separation
1
Press the front-panel
Preset
key.
2
Press
List by WL
.
3
Press the Delta
On
key.
Use the
Off
key to turn off the measurement.
4
Select the type of separation to observe:
•
∆
WL
displays channel separation.
•
∆
WL /
∆
PWR
displays both channel separation and differences in power.
5
Use the
and
softkeys to select the reference laser line.
6
Press
SELECT
.
Press
SELECT
at any time to select a new reference. Press
RESET
at any time to
turn off the delta calculation.
Measuring flatness
You can use relative power measurements to measure flatness (pre-emphasis)
in a WDM system. Simply select one carrier as the reference and measure the
remaining carriers relative to the reference level. The power differences rep-
resent the system flatness. Press
RESET
to turn off the delta calculations so
that all responses are shown in absolute wavelength and powers.
To measure flatness
1
Press the front-panel
Preset
key.
2
Press
List by Power
.
This lists the input signals by power with the largest response listed first.
3
Press the Delta
On
key.
4
Select
∆
PWR
.
5
Use the
and
softkeys to select the first laser line.
6
Press
SELECT
.
7
Since the largest power signal is the reference, the relative power
measurements for the other responses shows system flatness.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......