Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
–
[
dev
], the standard deviation of all power measurements measured.
On-Screen Messages
The messages listed in the table below may appear at the top of the screen during
the Stability application.
The PACT Application
What is the PACT ?
PACT (Passive Component Test Software) makes it possible to use your
Agilent 8163A Lightwave Multimeter or Agilent 8164A Lightwave
Measurement System to control a system for testing pigtailed or connectorized
passive devices (filters, couplers, and isolators) over wavelength. You must first
install a Tunable Laser module and Power Meters in your Agilent 8163A
Lightwave Multimeter or Agilent 8164A Lightwave Measurement System.
Depending on the configuration of the system you use, you can measure insertion
loss (single channel power measurement) over wavelength.
All results can be:
•
displayed on the screen,
•
printed out to hardcopy, or
•
saved to disk drive of your Agilent 8164A Lightwave Measurement System.
Table 5
Stability Application On-Screen Messages
Agilent 8163A
Message
Agilent 8164A
Message
Description
OvR
Overrange
One of the Power Meter channels has reported an over-
range. The over-ranged value is clipped, statistical data is
invalid.
NoH
No Head
There is an Optical Head Interface module inserted in the
mainframe that does not have an Optical Head connected to
it. No measurement is possible using this channel until you
either:
•
reconnect an Optical Head or
•
deselect this Optical Head channel as a power
measurement channel.
NoD
No Data
No data has been measured or measurement is not yet fin-
ished. You cannot access the Analysis Screen yet. Printout
and file operations are not yet possible.