Maintaining the PCM
18
Maintaining Your GC
331
Calibrating the PCM Interface
The interface's flow module contains a pressure sensor that
must be zeroed after it is installed on the GC. Calibration
ensures an accurate interface pressure display.
Do not connect the carrier gas to the flow module until you
have zeroed the interface's pressure sensor. For more
information on zeroing the pressure sensor, refer to the
1
If the gas supply is connected to the GC, turn off the
supply at the source, then disconnect the supply line from
the PCM inlet fitting.
2
Turn on the GC and wait 15 minutes to allow it to reach
thermal equilibrium.
3
When the GC has reached thermal equilibrium, press
[Options]
, scroll to Calibration and press
[Enter]
.
4
Scroll to the module to be zeroed and press
[Enter]
.
5
Scroll to a zero line and press
[Info]
. The GC will remind
you of the conditions necessary for zeroing that specific
sensor.
Flow sensors. Verify that the gas is connected and flowing
(turned on).
Pressure sensors. Disconnect the gas supply line at the
back of the GC. Turning it off is not adequate; the valve
may leak.
6
Press
[On/Yes]
to zero or
[Clear]
to cancel.
7
Turn off the GC.
8
Plumb the carrier gas to the flow module.
9
Turn on the GC.
If you were calibrating the flow sensor after replacing the
PCM, check for leaks.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...