Maintaining the G3435A/G3436A FPD
16
Maintaining Your GC
315
To Change the FPD Wavelength Filter
1
Gather the following:
•
Sulfur filter with filter spacer (See
•
Phosphorus filter
•
Cotton swab
•
Lens tissue
•
Lint- free gloves
2
Load the
and wait for the GC to
become ready.
3
Turn off the photomultiplier tube (PMT).
a
Select
[Front Det]
or
[Back Det]
.
b
Scroll to
PMT voltage
.
c
Press
Off
.
C A U T I O N
Do not touch the filter with your bare hands. For optimum
performance and to avoid scratches, use lint-free gloves for
assembling and inserting the filter into the assembly.
WA R N I N G
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
C A U T I O N
The photomultiplier tube (PMT) is extremely sensitive to light.
Always turn off the electrometer (which turns off the high voltage to
the PMT) before removing the PMT housing or opening the
emissions chamber. Failing to do this can destroy the PMT.
Even with the electrometer off, protect the PMT from room light.
Cap the housing after it is removed, place it end down to exclude
light, or reduce the room light level before exposing the PMT. A brief
exposure (always with the electrometer turned off) will not damage
it, but prolonged exposure will cause a gradual loss of sensitivity.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...