240
Maintaining Your GC
12
Maintaining the TCD
To Bakeout Contaminants from the TCD
The bakeout can be performed with the column installed or
the detector capped.
1
If the column is not installed, cap the detector.
2
Turn off the TCD filament.
3
If the column is attached to the inlet, maintain inert
carrier gas flow through the column.
4
Set the reference gas flow rate between 20 and
30 mL/min.
5
Set the detector temperature to 375 °C.
6
Hold at 375 °C for several hours.
7
If the column is uninstalled, install it. (See
8
Load the analytical method.
9
Allow the oven, inlet, and detector to equilibrate at
operating temperature, then re- tighten the fittings.
C A U T I O N
If the column is not installed, you must turn off the TCD filament and
cap the detector column fitting to prevent irreparable damage to the
filament caused by oxygen entering the detector.
WA R N I N G
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
WA R N I N G
If using hydrogen as a carrier gas, turn off the hydrogen supply
and cap the end of the column to prevent an oven explosion.
WA R N I N G
Be careful! The oven or detector fittings may be hot enough to
cause burns.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...