Maintaining Your GC
7
Maintaining the COC Inlet
To Replace a Needle in a Syringe
1
Gather the following (see
for injections onto 0.25- mm fused silica columns,” on
page 103):
•
Syringe barrel
•
Needle, 250- µm or 320- µm
2
Unscrew the syringe barrel cap and remove the spring.
3
Make sure that the needle has the PTFE disk as shown
below. If the syringe barrel does not have the PTFE disk,
use the instructions in the syringe box to wrap the
needle.
4
Check the new needle for a small wire inserted for
shipment. Remove the wire if present.
5
Slide the spring and the cap over the needle.
6
Insert the needle into the syringe barrel.
7
Screw the cap back on the syringe barrel.
Plunger
Barrel
Needle
Spring
Cap
PTFE disk
Needle
Stop
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...