2.2.1 Check condition of argon gas line
(1) Purge Ar line.
(2) Do leak check on Ar line.
(3) Check specification of Ar gas to have
≧
99.99% purity (In case cylinder has been replaced).
(4) Try alternative source of Ar gas (In case cylinder has been replaced and any alternatives
are available)
(5) Does this fix the problem?
Summary of Contents for Agilent 7700 Series ICP-MS
Page 15: ...2 4 6 Check torch box 6 Check torch stand Torch stand with clean contact Previous Next Good ...
Page 19: ...2 7 1 Fix torch box issues 1 Replace or maintain the parts 2 Did this fix the problem Yes No ...
Page 23: ...2 11 1 Check cooling water 1 Replace water with pure water 2 Did this fix the problem Yes No ...
Page 25: ...3 2 1 Check if photo sensor cable is damaged 1 Is the Cable for photo sensor damaged Yes No ...
Page 34: ...4 8 1 Check usage of User Tune 1 Is User Tune function in use Yes No ...
Page 40: ...6 2 1 Drain spray chamber 1 Pull out drain tube and drain liquid from the spray chamber Next ...
Page 50: ...9 1 End of procedure Close this window Return to top ...