68
Installation, Operation, and Maintenance
Part 1, Installation
2
Installation
Making A Trial Run
Once installation, configuration and updating are complete, make a quick
injection using the sampler to verify that it works properly.
1
Install an empty syringe in the injector.
2
If using a tray with a G2913A injector
, install the 1-sample turret.
If using a tray with a G2613A injector
, install the 3-sample turret.
3
Place empty bottles in all the Solvent A and Waste A turret positions. Place
an empty capped sample vial in the tray 1 position (or the turret sample 1
position, if not using the tray).
4
Make the sampler settings shown in
. These are designed for a
6890N GC. If you are using a different GC, use these settings as a guide. To
enter the settings:
5890 Series II GC
Use the G2911AA software.
For 6850 GC
Use the handheld controller.
For 6890 GCs
Use the GC keyboard.
Table 11
Trial run parameters
Parameter
Setting
Injection volume
1
#Sample pumps
1
Viscosity delay
0
#Sample washes
1
#Solv A washes
1
#Solv B washes
0
Slow plunger
Off
Pre-dwell time
0
Post-dwell time
0
Sampling offset
Off
#Solv A pre-wash
1
#Solv B pre-wash
0
#Injections/vial
1
Samples
1–1
(for 6890 GC with tray and Agilent ChemStation
control, enter 1 in vial box and run method).
Summary of Contents for agilent 7683B
Page 10: ...10 Installation Operation and Maintenance...
Page 18: ...18 Installation Operation and Maintenance...
Page 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Page 72: ...72 Installation Operation and Maintenance...
Page 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Page 202: ...202 Installation Operation and Maintenance A Cables and Connectors...