Part 1, Installation
Check your work
The injector must be vertical.
The alignment pin must be properly seated in the alignment hole.
The injector feet should touch the inlet cover.
If the injector will not sit upright on the GC, check that the plumbing and
cabling under the inlet cover are properly routed in their channels.
Choosing the turret type
The G2913A injector comes with two turrets, a 1-sample transfer turret for use
with or without a tray, and an 8-sample turret to be used without a tray. The
injector comes with the transfer turret installed.
If you choose not to use the tray, you may remove the 1-sample turret and
install the 8-sample turret. See
on page 147 for details.
If using a G2613A injector with a tray, install the 3-sample turret.
If using a G2913A injector with a tray, install the 1-sample turret.