213
Chapter 7 Plug-in Modules
N2270A 10-Channel High Voltage MUX Module
4
7
N2270A Wiring Information
The 28-pin connector on the rear panel is an AMP 1800V Plug Connector
205689-2.
An Agilent N2320A terminal block can be used to wire external high
voltage signals to the N2270A. Included in the N2320A are an AMP
1800V Receptacle Connector, 205690-2 and an AMP AWG24-20 Socket
66399-4. The connector has a metal strain relief and you must make sure
to use additional insulation around the wires to prevent the wire
insulation being cut, broken or otherwise damaged at the strain relief
clamp. Refer to page 283 for more information about the N2320A.
WARNING
Voltages greater than 30 Vrms, 42 Vpk or 60 Vdc present an electric shock
hazard. Disconnect source voltages before removing or connecting the
source-to-module I/O connector or wiring the connector. All field wiring
must be rated for the highest voltage applied to any channel.
N2270A Pinout The pinout assignments are shown below.
Column
1
Column
2
Column
3
Column
4
Column
5
Column
6
Column
7
Row1
CH2_H
CH3_H
CH4_H
Not Used
CH5_H
CH6_H
CH7_H
Row2
CH2_L
CH3_L
CH4_L
Not Used
CH5_L
CH6_L
CH7_L
Row3
CH0_H
CH1_H
Not Used
COM_H
Not Used
CH8_H
CH9_H
Row4
CH0_L
CH1_L
Not Used
COM_L
Not Used
CH8_L
CH9_L
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...