Chapter 2 Front-Panel Menu Operation
To Select the Output Termination
4
2
To Select the Output Termination
The Agilent 33500 Series has a fixed series output impedance of 50
to
the front-panel channel connectors. If the actual load impedance is
different than the value specified, the displayed amplitude and offset
levels will be incorrect. The load impedance setting is simply provided
as a convenience to ensure that the displayed voltage matches the
expected load.
1 Press
or
or
to open the channel configuration
screen. Note that the current output termination values (both
50
in this case) appear on the tabs at the top of the screen.
2 Specify the output termination.
Press the
Output Load
softkey.
3 Select the desired output termination.
Use the knob or numeric keypad to select the desired load impedance or
press the
Set to 50
softkey or the
Set to High Z
softkey.