148
Chapter 3 Features and Functions
Frequency Sweep
3
Marker Frequency
If desired, you can set the frequency at which the signal on the front-
panel
Sync
connector goes to a logic low during the sweep. The Sync
signal always goes from low to high at the beginning of the sweep.
• Marker frequency: 1
Hz to 30 MHz (limited to 200 kHz for ramp).
The default is 500 Hz
.
• When the sweep mode is enabled, the marker frequency
must
be
between the specified start frequency and stop frequency. If you
attempt to set the marker frequency to a frequency not in this range,
the waveform generator will automatically set the marker frequency
equal to the start frequency or stop frequency (whichever is closer).
• The Sync enable setting is overridden by enabling the marker used
with the sweep mode (
). Therefore, when the marker is
enabled (and the sweep mode is also enabled), the Sync setting is
ignored.
•
Front-Panel Operation
: After enabling a sweep, press
, then
Sync Setup
. Then press
Mode
and
Marker
. Finally, press
Marker
Freq
. Then use the knob or numeric keypad to enter the desired
marker frequency.
•
Remote Interface Operation
:
[SOURce[1|2]:]MARKer:FREQuency
{<
frequency
>|MINimum|MAXimum}
Summary of Contents for Agilent 33500 Series
Page 13: ...12 ...
Page 16: ...1 1 Quick Start ...
Page 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Page 42: ...2 2 Front Panel Menu Operation ...
Page 90: ...3 3 Features and Functions ...
Page 218: ...4 4 Application Programs ...
Page 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Page 244: ...5 5 Tutorial ...
Page 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Page 262: ...6 Specifications ...