MMEMory Subsystem
l
If a sequence file (.seq) is specified, all the arbitrary waveforms defined in the file are loaded.
l
If the waveform referenced by <
filename
> has already been loaded, the instrument will generate error number
+786, "Specified arb waveform already exists". Deleting an existing waveform requires clearing the waveform
non-volatile memory with
.
MMEMory:LOAD:LIST[1|2] <
filename
>
MMEMory:STORe:LIST[1|2] <
filename
>
Loads or stores a frequency list file (.lst).
Parameter
Typical Return
Any valid file name on the mass memory device
(none)
Store the current frequency list to a LIST file on internal storage:
MMEM:STOR:LIST "INT:\FreqList.lst"
Load a LIST file on the internal storage mass memory system (file named FreqList.lst).
MMEM:LOAD:LIST "INT:\FreqList.lst"
l
A frequency list controls frequency in
, rapidly changing to the next frequency in the list when a
trigger event is received.
l
A frequency list file contains a comma-separated sequence of ASCII numbers, with the first number representing
the number of frequencies in the list:
3, 100.000000, 1000.000000, 550.000000
l
If the destination file exists, it is overwritten, unless marked as hidden or system.
MMEMory:LOAD:STATe <
filename
>
MMEMory:STORe:STATe <
filename
>
Stores the current instrument state to a state file. The file name optionally includes the folder name and the .sta file
extension.
Parameter
Typical Return
Any valid file name on the current directory
(none)
Store the current instrument state to the state file MyFreqMeas.sta in the root directory of the internal flash file sys-
tem.
MMEM:STOR:STAT "INT:\MySetup"
Load the instrument state from MySetup.sta in the root directory of the internal storage.
MMEM:LOAD:STAT "INT:\MySetup.sta"
MMEMory:MOVE <
file1
>,<
file2
>
Moves and/or renames <
file1
> to <
file2
>. The file names must include the file extension.
Agilent 33500 Series Operating and Service Guide
257