1260 Infinity Binary LC - System User Guide
5
Quick Start Guide
Data Analysis
Integrating a Signal
1
Select the Integration Task Tool (see figure below). The
Integrate
icon
and the
Set Integration Events Table
icon are highlighted in the figure
shown below.
2
Click the
Set Integration Events Table
icon
to open the table as shown.
3
Set
Baseline Correction
to
Advanced
for gradient runs.
4
Set
Slope Sensitivity
to 5.
Higher numbers will integrate steeper peaks and ignore less steep
peaks.
5
Set the
Peak Width
value to the narrowest peak of interest,
approximately 0.02 in this case.
6 Area Reject
and
Height Reject
can be set to reject the smallest peaks.
7
Click the
Integrate
icon
to update the results using these new
settings.
8
Exit from the
Integration Events
table using the green tick icon
.