1260 Infinity Binary LC - System User Guide
5
Quick Start Guide
Data Acquisition in Method and Run Control View
Running the Method for a Single Injection
This section shows how to run a single injection of the isocratic checkout
sample using the conditions entered in the previous section.
Analyses on the ChemStation can be run in two modes:
• Run Method
— single injections, for example, in interactive method
development, using the parameter settings current at the time;
• Run Sequence
— automated series of injections from multiple vials,
possibly with multiple methods. For further details refer to the
ChemStation manuals.
1
Click the
Select Run Method Task
icon
.
2
If required method conditions are not currently loaded, select
Method >
Load Method
or the icon
under the menu bar to load them.
3
Place the sample vial in position 1 — this is the front position of the 10
x 2 ml vial positions on the right hand side of the sample tray.
4
Select
Run Control > Sample Info
, and enter the required sample
parameters.
For example:
•
Subdirectory
(optional)
•
Name Pattern
•
Vial/Location
•
Sample name
•
Comment
5
If the system is already equilibrated and the baseline is stable, click
Run
Method
in the
Sample Info
page to start the injection. Alternatively, click
OK
, and when ready, select
Run Control > Run Method
. The item will be
added to the
Run Queue
and then automatically started from there.
N O T E
If changes have been made to a loaded method and not yet saved, this is indicated by a
yellow asterisk on the method status icon. The injection can be made without first saving
those parameters changes. The ChemStation always stores a copy of the acquisition
parameters in the data file as ACQ.TXT to ensure the preservation of the original method
parameters.