16
Instructions for Use
Error Messages
Error Messages
For a certain type of failure, the controller will self-diagnose the
error and the messages described in the following table are
displayed.
Tab. 2
MESSAGE DESCRIPTION
REPAIR
ACTION
CHECK CONNECTION TO
PUMP
Wrong connection between the pump
and the controller.
Check connection between
controller and pump.
Press the START push-button
twice to start the pump.
PUMP WAITING INTERLOCK
The interlock signal of P1 connector is
activated by an interruption of the link
between pin 3 and 8 of J1 connector,
or because the external interlock signal
is open.
Reset the short circuit between
pin 3 and pin 8 of J1 connector,
or close the external interlock
signal.
FAULT: PUMP OVERTEMP.
The upper bearing/pump temperature
exceeds 60 °C.
The lower bearing/pump temperature
exceeds 80 °C.
Wait until the temperature
decrease below threshold value.
Press the START push-button
twice
to start the pump.
FAULT: CONTROLLER
OVERTEMPERATURE
The controller transformer temperature
exceeds 90 °C or the temperature on
the outgoing Mosfets radiator exceeds
60
°
C.
Wait until the temperature
decrease below threshold value.
Press the START push-button
twice
to start the pump.
234/306
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O
)
Summary of Contents for 969-9462
Page 4: ...Turbo V 2000HT Controller 4 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 124: ...8 Bruksanvisning Felmeddelanden 124 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 152: ...10 Ohjekäsikirja Vianetsintä 152 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 208: ...14 Návod na Obsluhu Oznamy vád 208 306 Turbo V 2000HT Controller User Manual 87 900 913 01 O ...
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......