Technical Information
Procedure to Connect the Serial and I/O ports
to an External Cable
282/306
Turbo-V 903 Pumping System User Manual
/ 87-901-016-01
(A)
Procedure to Connect the Serial and I/O ports to an External
Cable
In the following picture, it has showed the right procedure to connect a cable
to the I/O or to the serial port.
A shielded cable of 3 m maximum length has to be utilized for both serial and I/O
port connection. Remove the plastic protection on both the ends of the cable
(picture b) and after it is necessary to connect the wires to the single pins like
showed in the picture c.
NOTE
Take care to have a good contact between the metallic case and the external
shield of the wire. Moreover, this operation has to be done at least on the
controller side.
In this way, you will be sure to reduce the influence of the external noise and
to accomplish the EMC requests. In picture d is showed the cable assembled.
Figure 19
Cable assembled
Summary of Contents for 969-6500
Page 4: ...Turbo V 903 Pumping System 4 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 14: ...Contents 14 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 58: ...3 Mode d emploi Mise au rebut 58 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 118: ...7 Istruktionsbog Bortskaffelse 118 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 176: ...11 Oδηγιεσ Χρησεωσ Χώνευση 176 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 218: ...14 Návod k Použití Likvidace 218 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 232: ...15 Návod na Obsluhu Likvidácia 232 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......