Technical Information
Vaclon Plus Pump Replacement Parts and Accessories
5
VacIon Plus 300 pumps User Manual
/
87-900-103-01 (H)
91/98
Symptom
Possible cause
Correction procedure
7) – System fails to achieve desired
UHV vacuum pressure.
System not fully baked. Water
vapor limits base pressure.
Bake the system walls,
thereby accelerating the
desorption process. Baking
mobilizes the vapours so
they can be cracked and
pumped by discharge (see
section “BAKEOUT
OPERATION”).
8) – System fails to achieve desired
UHV vacuum pressure.
System not appropriately cleaned
for UHV. Excessive outgassing from
walls limits base pressure.
Clean all components for
UHV and bake the system
again.
Vaclon Plus Pump Replacement Parts and Accessories
Tab. 3
Part Number
Diode
Noble Diode
Starcell
Basic pump
919-1610
919-1620
919-1640
Pump with options
With heaters installed
(120 V)
(220 V)
919-1611
919-1612
919-1621
919-1622
919-1641
919-1642
Double-ended
919-1650
919-1660
919-1680
Double-ended with
heaters (120 V)
installed (220 V)
919-1651
919-1652
919-1661
919-1662
919-1681
919-1682
Cables and accessories
HV rad. Resist. cable,
with interlock
929-0705
Summary of Contents for 9191610
Page 3: ...VacIon Plus 300 VacIon Plus 300 pumps User Manual 87 900 103 01 H 3 98 VacIon Plus 300...
Page 4: ...VacIon Plus 300 4 98 VacIon Plus 300 pumps User Manual 87 900 103 01 H...
Page 8: ...Contents 8 98 VacIon Plus 300 pumps User Manual 87 900 103 01 H...
Page 56: ...4 Installation Procedure Disposal 56 98 VacIon Plus 300 pumps User Manual 87 900 103 01 H...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......