![Agilent Technologies 8960 Series 10 Reference Manual Download Page 420](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627420.webp)
420
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_orfspectrum.fm
SETup:ORFSpectrum
SETup:ORFSpectrum:TIMeout:STATe
SETup:ORFSpectrum:TRIGer:DELay
SETup:ORFSpectrum:TRIGger:SOURce
Function
Selects/queries output RF spectrum measurement timeout state.
Setting
0|OFF | 1|ON
Query
0|1
*RST Setting
0|off
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:TIMEOUT:STATE ON" !Sets timeout state to on.
Function
Sets/queries the trigger delay for output RF spectrum measurements. The units (S|MS|US|NS)
are optional, if no units are specified than units default to S.
Range
Range:
−
2.31 ms to +2.31 ms
Resolution: 5 significant digits or 100 nanoseconds whichever is greater
Query
Range:
−
2.31 ms to +2.31 ms
Resolution: 5 significant digits or 100 nanoseconds whichever is greater
*RST Setting
0 seconds
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:TRIGGER:DELAY 1MS" !Sets the trigger delay value
!to 1 millisecond.
Function
Selects/queries the trigger source for output RF spectrum measurements. See
Spectrum Measurement Description” on page 75
Setting
Range: AUTO | IMMediate | PROTocol | RISE
See
“Triggering of Measurements” on page 149
Query
Range: AUTO | IMM | PROT | RISE
*RST Setting
AUTO
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:TRIGGER:SOURCE AUTO" !Sets trigger source.
Summary of Contents for 8960 Series 10