404
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_dynp.fm
SETup:DPOWer
SETup:DPOWer:COUNt:NUMBer
SETup:DPOWer:EMDifference
Function
Sets/queries the number of bursts for the Dynamic Power measurement.
Setting
Range: 1 to 100
Resolution: 1
Query
Range: 1 to 100
Resolution: 1
*RST
10
Programming Example
OUTPUT 714;"SETUP:DPOWER:COUNT:NUMBER 30" !Sets the number of bursts for
the Dynamic Power measurement to 30.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Sets/queries the Expected Maximum Difference from Previous Measurement parameter for the
Dynamic Power measurement. The units dB are optional.
The Expected Maximum Difference from Previous Measurement parameter is used with the
measured transmit power from the previous burst to set the maximum RF power that the base
station emulator is expecting the mobile to transmit in the next burst.
The setting of this parameter does not affect the receiver Expected Power parameter. See
“RFANalyzer:EXPected:POWer[:SELected]” on page 372
Setting
Range: -30 dB to +30 dB
Resolution: 0.01 dB
Query
Range: -30 dB to +30 dB
Resolution: 0.01 dB
Programming Example
OUTPUT 714;"SETUP:DPOWER:EMDIFFERENCE -3" !Sets the Expected Maximum
! Difference from previous measurement parameter
! to -3 dB. (Example: If the previous burst
! measures 0 dB and you expect the maximum
! amplitude of the next burst to be 3 dB, set the
! Expected Maximum Difference parameter to 3).
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX