17 Maintaining a Valve
To Align a Rotary Valve Rotor
270
Maintaining Your GC
To Align a Rotary Valve Rotor
1
Gather the following:
•
Flathead screwdriver
•
3-mm hex key wrench
•
T-20 Torx screwdriver
2
Set the oven and valve box heated zones to a safe handling temperature (25 °C).
3
Set all valves to
Off
.
WARNING
The oven, inlet, detector, and valve box may be very hot. If they are hot, wear heat-resistant
gloves to protect your hands.
4
Loosen the adjustment set screw.
5
Locate the rotor adjustment shaft on top of the actuator. Using a flathead screwdriver,
rotate the valve rotor counterclockwise until it stops, then back it off a small amount to set
one end of the rotor’s motion (< 1 mm).
6
Tighten the adjustment set screw.
7
Turn the valve
On
,
turn
Off
to check for smooth operation.
8
Restore the analytical method.
Rotor adjustment
slotted shaft
Adjustment set screw
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...