10 Maintaining the VI
To Remove the VI Interface
Maintaining Your GC
171
To Remove the VI Interface
1
Gather the following:
•
1/4-inch and 7-mm wrench
•
T-20 Torx screwdriver
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The interface may be hot enough to cause burns. If the interface is hot, wear
gloves to protect your hands.
3
Remove the column.
4
Remove the transfer line by loosening the nut, then lift it from the interface.
5
Loosen the five inlet cover screws and remove the inlet cover.
6
Remove the clamping plate from the interface by loosening the captive screw.
7
Lift the interface out of the heater block.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...