![Agilent Technologies 8890 Maintaining Manual Download Page 158](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678158.webp)
9
Maintaining the PTV Inlet
To Replace the Inlet Adapter for the PTV Inlet
158
Maintaining Your GC
To Replace the Inlet Adapter for the PTV Inlet
1
Select from the following list an adapter with the smallest hole diameter that will accept
the column. The adapter number is stamped on the side of the adapter. Also see
“Consumables and Parts for the Programmable Temperature Vaporization Inlet”
2
Gather the following:
•
Replacement adapter
•
Replacement silver seal
•
6-mm wrench
•
5-mm wrench
•
Lint-free gloves
3
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4
Unscrew the column nut from the adapter. Remove the nut and the column from the inlet.
Table 26 PTV inlet adapters
Column id
Inlet adapter number
*
* Includes (1) adapter, (1) silver seal, and (1) split column nut.
Part number
0.20 mm
31
5182-9754
0.25 to 0.33 mm
45
5182-9761
0.53 mm
70
5182-9762
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...