![Agilent Technologies 8890 Maintaining Manual Download Page 152](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678152.webp)
9
Maintaining the PTV Inlet
To Replace the Septumless Head PTFE Ferrule on the PTV Inlet
152
Maintaining Your GC
To Replace the Septumless Head PTFE Ferrule on
the PTV Inlet
1
Gather the following:
•
Syringe with 23-gauge needle, see
“Consumables and Parts for the Programmable
Temperature Vaporization Inlet”
•
Replacement PTFE ferrule
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
3
Unscrew the guide cap from the septumless head and remove the PTFE ferrule.
4
Push the guide cap and replacement ferrule over the syringe needle so that at least 10 mm
of the needle tip is exposed.
5
Guide the end of the needle into the septumless head until the ferrule meets the
septumless head.
6
Loosely install the guide cap.
7
If not configured, configure the column.
8
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
9
Put the inlet into
Splitless
mode.
10
Set the column flow to 5 mL/min and the purge flow to 60 mL/min.
11
After the inlet pressurizes, press
once.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...