![Agilent Technologies 8890 Maintaining Manual Download Page 129](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678129.webp)
8
Maintaining the MMI
To Install a Capillary Column with the Multimode Inlet
Maintaining Your GC
129
Figure 29.
Insert Smart ID key
14
Configure the new column.
15
Condition the column per the manufacturer’s recommendation. See
16
Install the column into the detector.
•
To Install a Capillary Column in the FID
•
To Install a Capillary Column in the NPD
•
To Install a Capillary Column in the TCD
•
To Install a Capillary Column in the ECD
•
To Install a Capillary Column to the FPD+
17
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
18
After the column is installed at both inlet and detector, establish a flow of carrier gas and
purge as recommended by the column manufacturer.
19
Restore the analytical method.
•
For FID or FPD+, immediately turn off the flame.
•
For NPD, immediately turn off the bead.
20
After the GC becomes ready, wait 10 minutes then ignite the detector flame or bead.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If they are hot,
wear heat-resistant gloves to protect your hands.
21
Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
the fittings.
22
Perform an inlet
Leak & Restriction
test and reset the maintenance counters.
Smart ID key
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...