Maintaining the COC Inlet
To Replace the 7693A Injector Needle Support Guide
116
Maintaining Your GC
To Replace the 7693A Injector Needle Support
Guide
Before using a 7693A ALS to make injections onto the COC inlet, install the on-column needle
support guide.
Refer to the 7693A ALS documentation for complete details.
1
Open the injector door.
2
Remove the syringe.
3
Slide the syringe carriage up to the top position.
4
Completely remove the T-10 Torx screw from the support foot. Be careful to not let the
screw fall into the turret assembly.
5
Slide off the support foot.
6
Slide on the new support foot.
7
Replace the T-10 Torx screw and tighten.
8
Install the appropriate syringe.
9
Close the injector door.
10
Align the injector.
Standard needle
support guide
On-column
needle support
guide
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...