Auto Align
Auto Align
Pressing the auto align button on the front panel of the instrument performs
an automatic alignment of the instrument using the largest signal found in a
full span sweep. This aligns the output of the monochromator with the photo-
detector for improved amplitude accuracy. To ensure maximum amplitude
accuracy and stability, connect an input signal to the instrument, mark the
peak of the signal, and then press Auto Align. This starts an automatic align-
ment procedure that should be performed whenever the instrument has been:
• moved
• subjected to a 2
°
temperature change
• restarted and warmed up at the start of each day
The automatic alignment requires the connection of an external, broadband or
narrowband, light source. If there is insufficient signal power, the automatic
alignment will not be performed and an error message will be reported.
The auto align function saves and restores the current instrument state. This
allows the auto align function to be used in the middle of a measurement rou-
tine.
If markers are turned on, the auto align function attempts to do the automatic
alignment at the wavelength of the active marker. If the instrument is in zero
span, the alignment is performed at the center wavelength.
Key Path
Auto Align
Related Functions
Auto Meas
Auto Align & Add to Trajectory
Calibrator Multi-Pt Align
User Source Multi-Pt Align
Remote
Commands
CALibration:ALIGn
CALibration:ALIGn:MARKer[1|2|3|4]
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...