3-102
Function Reference
Wavelength Line Mkr 1/2
Signal Source
Selects either an external single wavelength signal source or the internal cali-
brator as the wavelength calibration source.
Wavelength Referenced In (for external calibration source only)
Selects air or vacuum for the calibration. If the internal calibrator is selected,
this selection will not be available.
Set Calibration Wavelength (for external calibration source only)
This is the exact wavelength of the calibration source that will be used for the
next user wavelength calibration. The wavelength entered must be within
2.5 nm of the wavelength measured using the factory calibration. If the inter-
nal calibrator is selected, this selection will not be available.
User Multipoint Wavelength Calibration Data
Selects either Offset or Replaced for the calibration data. Offset will adjust the
user multipoint data at all wavelengths to provide the best wavelength accu-
racy at the calibration source wavelength.
Replaced will disable the multipoint data when used with an external source
(select Offset to re-enable).
Replaced will delete multipoint data when used with the internal calibrator.
Key Path
System > Calibration > Wavelength Cal Setup
Remote
Commands
CALibrate:WAVelength:STATe ON
CALibrate:WAVelength:STATe OFF
CALibrate:WAVelength:INTernal
Wavelength Line Mkr 1/2
Allows you to set the positions of the line markers. Wavelength Line Mkr 1 is
always to the left of Wavelength Line Mkr 2. When either one of the line mark-
ers is accessed, the reduced section of the wavelength is highlighted in blue.
The line marker can then be set to the desired position using the knob, step
keys, or numeric keypad. This limits the measurement range of the sweep
limit, search limit, and integrated power.
Key Path
Markers > More Marker Functions > Line Marker Menu > Wavelength Line
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...