CALCulate3 Subsystem
5
-
33
:CALCulate3:DELTa:PRESet
Turns off all delta measurement states.
Attribute Summary
Preset
State: not affected
*RST State: not affected
SCPI Compliance: instrument specific
Command Only
:CALCulate3:DELTa:REFerence:FREQuency{?| {<real> | MINimum | MAXimum}}
Selects the reference laser line for DELTa calculations.
<real>
is a frequency value that is within the following
limits:
Attribute Summary
Preset
State: 236.0571 THz (1270 nm)
*RST State: 236.0571 THz (1270 nm)
SCPI Compliance: instrument specific
Description
The reference will be the laser line at the frequency
closest to the frequency entered. Subsequent measure-
ments will use the frequency closest to the reference
frequency used for the previous measurement.
The query returns the reference laser line’s frequency.
The default units for the
<real>
parameter are Hz.
:CALCulate3:DELTa:REFerence:POWer?
Queries the reference laser line’s power level.
Attribute Summary
Preset
State: not affected
*RST State: not affected
SCPI Compliance: instrument specific
Query Only
:CALCulate3:DELTa:REFerence[:WAVelength]{?| {<real> | MINimum | MAXimum}}
Selects the reference laser line for DELTa calculations.
<real>
is a wavelength value that is within the following
limits:
Constant
Description
MINimum
181.6924 THz
MAXimum
236.0571 THz
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......