5
-
10
Common Commands
Example
The following example starts the data acquisition
according to the current settings.
OUTPUT 720;”*TRG”
*TST?
The *TST (test) query starts a self-test on the instru-
ment.
Description
The result of the test is placed in the output queue. A
zero indicates the test passed and a non-zero value indi-
cates the test failed. The instrument will initiate a mea-
surement and check for any hardware errors after the
measurement is complete.
Query Response
<integer>
Example
OUTPUT 720;”*TST?”
ENTER 720;Result
PRINT Result
*WAI
The *WAI command prevents the instrument from exe-
cuting any further commands until the current com-
mand has finished executing.
Description
All pending operations are completed during the wait
period. This command cannot be issued as a query.