Common Commands
5
-
7
results in error number –230,
“Data corrupt or stale”
.
You must initiate a new sweep with :INIT:IMM before
you can use the :FETCh command.
*SAV <integer>
This command saves an instrument state.
<integer>
range is 1 to 4.
Item
Setting
Start wavelength
1270 nm
Stop wavelength
1650 nm
Graphical display
off
Measurement acquisition
single
Elevation correction value
0 meters
Wavelength units
nm
Power offset
0 dB
Peak threshold
10 dB
Peak excursion
15 dB
Measurement speed
normal
Number of uncorrected data points
15,047
Device bandwidth
narrowband
Delta Measurements:
∆
power
off
∆
wavelength
off
∆
wavelength and power
off
reference signal position
1270 nm
Drift measurements
off
Signal-to-Noise Measurements:
measurement
off
number of averages (count)
100
GPIB address
not affected
Serial port settings
not affected
Power-bar display
on
Selected printer
internal
Remote interface setting
not affected
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......