22-3
TDR/TDT Commands (Rev. A.06.00 and Above)
Module Channel
Identification
In previous software revisions, each TDR/TDT subsystem command identified the TDR mod-
ule installation (left or right mainframe slot) with the form
:TDR{2:4}:<command>
. Starting with
software revision A.06.00, the TDR/TDT subsystem no longer uses this identification scheme;
the new syntax form is simply
:TDR:<command>
.
TDR/TDT
Calibration
TDR/TDT calibration corrects for measurement system effects and locates the reference
plane of the step response. The reference plane is the time (or distance) of the incident step
and is the location that all subsequent impedance measurements are made relative to. Start-
ing with software revision A.06.00 and above, TDR/TDT Calibration replaces the normaliza-
tion and reference plane calibration. TDR/TDT Calibration allows marker time readouts
relative to the reference plane but, in addition, adds the ability to change the time base set-
ting, corrects for systematic errors, and enables a pulse rise time filter to simulate real step
responses. For best results, before starting the TDR/TDT calibration place the step response
at the reference plane within the first graticule division as shown the following picture.
The calibration commands step through the TDR/TDT Calibration Wizard. Send
“RESPonse:CALibrate” on page 22-5
to begin the calibra-
tion. Use the returned string from the SDONe? query to determine when a calibration step
has completed. If you set a time out value, make sure that the value is set long enough to
allow the measurement to complete. SDONe? returns the prompt string for the next step.
After making the test setup connections for a calibration step, send
followed by SDONe?. At the end of the last step, SDONe? returns the string “Done”.
N O T E
Once the module calibration procedure is started, all access to the instrument’s front panel is blocked, including
the use of the Local button. Pressing Local during a module calibration will not place the instrument in local
mode. The calibration must either be cancelled (using
) or finished before you can regain
control to the instrument’s front panel. Failure of a calibration step results in that step being repeated.
More Information
Option 202 TDR Peeling is implemented as a math function.
. To perform the measurements that are listed on the measurement toolbar, refer to
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......