19-6
S-Parameter Commands
MARKer:Y2Position?
MARKer:Y2Position?
Command
:SPARameter:MARKer:Y2Position?
Queries the amplitude value (Y2) of the X2 marker.
Query
:SPARameter:MARKer:Y2Position?
Returned Format
[:SPARameter:MARKer:Y2Position] <value><NL>
MARKer:XDELta?
Command
:SPARameter:MARKer:XDELta?
Queries the frequency difference (
Δ
) between the X1 and X2 markers.
Query
:SPARameter:MARKer:XDELta?
Returned Format
[:SPARameter:MARKer:XDELta] <value><NL>
MARKer:YDELta?
Command
:SPARameter:MARKer:YDELta?
Queries the amplitude difference (
Δ
) between the X1 and X2 markers (Y1 and Y2 positions).
Query
:SPARameter:MARKer:YDELta?
Returned Format
[:SPARameter:MARKer:YDELta] <value><NL>
VWINdow
Command
:SPARameter:VWINdow {ON | 1 | OFF | 0}
Turns on and off the display of the windowed time-domain region. This region highlights the
the range of the TDR data that will be transformed to the frequency domain and displayed on
the S-parameter graph. It is a visual aid for the user and does not alter the data range trans-
formed.
Example
10 OUTPUT 707;":SPAR:VWIN ON"
Query
:SPARameter:VWINdow?
Returned Format
[:SPARameter:VWINdow] {ON | 1 | OFF | 0}<NL>
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......