
18-41
Measure Commands
SINTegrity:BERFloor?
See Also
Refer to the MEASure:RESults query for information on the results returned and how they
are affected by the SENDvalid command. Refer to the individual measurements for informa-
tion on how the result state is returned.
SINTegrity:BERFloor?
Query
MEASure:SINTegrity:BERFloor?
Returns the bit error ratio (BER) floor measurement. This is the extrapolated BER at the
center of the eye. If both amplitude and jitter analysis is active, it will take into account both
the probability of timing errors (jitter) as well as the probability of amplitude errors (noise
and interference).
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Returned Format
[:MEASure:SINTegrity:BERFloor] <measurement><NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:BERFLOOR?”
SINTegrity:BERLimit?
Query
MEASure:SINTegrity:BERLimit?
Returns "JITTER" if the signal bit error ratio is primarily due to jitter, "AMPLITUDE" if the
signal bit error ratio is primarily due to noise and interference, or "BALANCED" if the two are
evenly contributing to BER. The value 9.999E37 is returned, if both BER Floors are
≤
1e-18.
This command is only available if both jitter analysis and amplitude analysis are turned on.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Returned Format
[:MEASure:SINTegrity:BERLimit] {JITTer | AMPLitude | BALanced}<NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:BERLIMIT?”
SINTegrity:PATTern?
Query
MEASure:SINTegrity:PATTern?
Returns definite-length block data. The data block contains the pattern as determined by the
instrument. Each value in the pattern is a single byte. Values in the pattern are the ASCII val-
ues for '0' and '1' (30 hex and 31 hex, respectively). The data block is followed by a termina-
tor character, 0A hex (linefeed). This query will return an incomplete description of the
pattern if all of the data needed to determine the pattern has not yet been acquired. This
query produces an error if signal type is set to clock signal. This replaces the obsolete com-
mand
“JITTer:PATTern?” on page 18-31
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Returned Format
[:MEASure:SINTegrity:PATTern] <value><NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:PATTERN?”
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......