16-6
Marker Commands
Y2Position
Returned Format
[:MARKer:Y1Position] <Y1_position><NL>
Example
This example sets the Y1 marker to 10 mV.
10 OUTPUT 707;":MARKER:Y1POSITION 10E-3"
Y2Position
Command
:MARKer:Y2Position <Y2_position>
This command sets the Y2 manual marker position and moves the Y2 manual marker to the
specified value on the specified source if the Y2 marker is in manual state. <Y2_position> is
the current measurement unit value at Y2.
Query
:MARKer:Y2Position?
The query returns the current measurement unit level at the Y2 marker position.
Returned Format
[:MARKer:Y2Position] <Y2_position><NL>
Examples
This example sets the Y2 marker to –100 mV.
10 OUTPUT 707;":MARKER:Y2POSITION -100E-3"
YDELta?
Query
:MARKer:YDELta?
This query returns the current measurement unit difference between Y1 and Y2 if they are
both on and both have the same source. If not, 9.37 is returned.
Vdelta = value at Y2 – value at Y1
Returned Format
[:MARKer:YDELta] <value><NL>
<value>
Measurement unit difference between Y1 and Y2.
Example
10 OUTPUT 707;":SYSTEM:HEADER OFF"
20 OUTPUT 707;":MARKER:YDELTA?"
YUNits
Command
:MARKer:YUNits {VOLT | OHM | REFLect}
This command sets the units for vertical display in TDR and TDT applications. The units may
be in volts, ohms, or % reflection. The marker mode must be TDRTDT to use this feature.
Query
:MARKer:YUNits?
This query returns the current marker units setting.
Returned Format
[:MARKer:YUNits]{VOLT | OHM | REFLect}<NL>
Example
10 OUTPUT 707;":MARKER:YUNITS OHM"
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......