
1-17
Introduction
Status Reporting
Table 1-5. Status Reporting Bit Definition (1 of 2)
Bit
Description
Definition
ACQ
Acquisition
Indicates that acquisition test has completed in the Acquisition Register.
AREQD
Autoscale Required
Indicates that a parameter change in Jitter Mode has made an autoscale necessary.
CLCK
CloCk
Indicates that one of the enabled conditions in the Clock Recovery Register has
occurred.
CME
Command Error
Indicates if the parser detected an error.
COMP
Complete
Indicates the specified test has completed.
DDE
Device Dependent Error
Indicates if the device was unable to complete an operation for device dependent
reasons.
EFAIL
Edge Characterization
Fail
Indicates that the characterizing of edges in Jitter Mode has failed.
ESB
Event Status Bit
Indicates if any of the enabled conditions in the Standard Event Status Register have
occurred.
EXE
Execution Error
Indicates if a parameter was out of range or was inconsistent with the current
settings.
FAIL
Fail
Indicates the specified test has failed.
JLOSS
Pattern Synchronization
Loss
Indicates that the pattern synchronization is lost in Jitter Mode.
LCL
Local
Indicates if a remote-to-local transition occurs.
LOCK
LOCKed
Indicates that a locked or trigger capture condition has occurred in the Clock Recovery
Module.
LOSS
Time Reference Loss
Indicates the Precision Timebase (provided by the Agilent 86107A module) has
detected a time reference loss due to a change in the reference clock signal.
LTEST
Limit Test
Indicates that one of the enabled conditions in the Limit Test Register has occurred.
MAV
Message Available
Indicates if there is a response in the output queue.
MSG
Message
Indicates if an advisory has been displayed.
MSS
Master Summary Status
Indicates if a device has a reason for requesting service.
MTEST
Mask Test
Indicates that one of the enabled conditions in the Mask Test Register has occurred.
NSPR1
No Signal Present
Receiver 1
Indicates that the Clock Recovery Module has detected the loss of an optical signal on
receiver one.
NSPR2
No Signal Present
Receiver 2
Indicates that the Clock Recovery Module has detected the loss of an optical signal on
receiver two.
OPC
Operation Complete
Indicates if the device has completed all pending operations.
OPER
Operation Status
Register
Indicates if any of the enabled conditions in the Operation Status Register have
occurred.
PON
Power On
Indicates power is turned on.
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......