7
The 85103F Upgrade Package (8510A/B to 8510C)
Checking 8510 System Operation
Checking 8510 System Operation
Use this procedure to ensure that the 8510A system works properly.
Procedure
1. Configure the system as shown in
2. Turn on the instruments in the following order:
a. Source (be sure to turn the switch to power on, not standby)
b. Test set
c. Analyzer
3. On the 85102, press the green preset key.
4. Does the system pass self-test (85101 self tests, 85102 running error messages, and test
set unratio power tests) and display a log MAG S
11
trace? If not, do not perform the
upgrade until the system has been serviced.
5. Initialize a blank or used tape (customer supplied), as follows:
a. Load the tape (with write-protect off).
b. On the 85102, press
AUXILIARY MENUS TAPE
.
c. On the 85101, select
INITIALIZE TAPE YES.
6. Does the system initialize the tape (this takes approximately three minutes) and
display
TAPE INITIALIZED
?
NOTE
The tape drive must be working to transfer information from tape to
disk.
7. If the system displays a self-test,
1
running error message, or unratio power test failure,
check the equipment setup. Do not perform the upgrade until either you correct the
setup problem, or you have the system serviced.
8. Continue with
“Converting Files from Tape to Disk.”