![Agilent Technologies 85054D User'S And Service Manual Download Page 36](http://html1.mh-extra.com/html/agilent-technologies/85054d/85054d_users-and-service-manual_2868611036.webp)
4-2
85054D
Performance Verification
Introduction
Introduction
The performance of your calibration kit can only be verified by returning the kit to Agilent
Technologies for recertification. The equipment required to verify the specifications of the
devices in the kit has been specially manufactured and is not commercially available.
How Agilent Verifies the Devices in This Kit
Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision
airlines and shorts that are directly traced to NIST (National Institute of Standards
and Technology). The airline and short characteristics are developed from mechanical
measurements. The mechanical measurements and material properties are carefully
modeled to give very accurate electrical representation. The mechanical measurements
are then traced to NIST through various plug and ring gages and other mechanical
measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale)
testing of the calibration devices, Agilent includes the test measurement uncertainty as
a guardband to guarantee each device meets the published specification. For
recertifications (after sale), no guardband is used and the measured data is compared
directly with the specification to determine the pass or fail status. The measurement
uncertainty for each device is, however, recorded in the calibration report that
accompanies recertified kits.
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the
institute’s calibration facility. The specifications data provided for the devices in this kit is
traceable to NIST through Agilent Technologies.
Summary of Contents for 85054D
Page 5: ...85054D 1 1 1 General Information...
Page 14: ...1 10 85054D General Information When to Calibrate...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85054D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 35: ...4 1 4 Performance Verification...
Page 38: ...4 4 85054D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85054D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85054D Replaceable Parts Introduction...
Page 49: ...A 1 A Standard Definitions...