![Agilent Technologies 85051B User'S And Service Manual Download Page 26](http://html.mh-extra.com/html/agilent-technologies/85051b/85051b_users-and-service-manual_2867929026.webp)
3- 8
85051B
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Procedure
NOTE
Always hold a connector gage by the gage barrel, below the dial indicator. This gives the best
stability, and improves measurement accuracy.
1. Select the proper gage for your connector. Refer to
for gage part number.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to
“Cleaning Connectors” on page 3-4
of this chapter.
3. Zero the connector gage (refer to
):
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage
master to the gage. Refer to
“Making Connections” on page 3-12
for more information. Connect the
nut finger tight. Do not overtighten.
b. Using an open-end wrench to keep the gage from rotating, use the torque wrench recommended for
use with the kit to tighten the gage master connector nut to the specified torque. Refer to
Connection Using a Torque Wrench” on page 3-12
for more information.
c. The gage pointer should line up exactly with the zero mark on the gage. If not, loosen the dial lock
screw on the gage and rotate the gage dial so that the pointer is aligned with the zero mark.
d. Tighten the dial lock screw and remove the gage master.
e. Attach and torque the gage master to the gage once again to verify that the setting is repeatable.
Remove the gage master.
4. Gage the device connector (refer to
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage
master to the gage. Refer to
“Making Connections” on page 3-12
for more information. Connect the
nut finger-tight. Do not overtighten.
b. Using an open-end wrench to keep the gage from rotating, use the torque wrench recommended for
use with the kit to tighten the device connector nut to the specified torque. Refer to
Connection Using a Torque Wrench” on page 3-12
for more information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. If the needle has moved clockwise, the center conductor is
protruding
by an amount indicated by the
black
numbers. If the needle has moved counterclockwise, the center
conductor is
recessed
by an amount indicated by the
red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take an average of the
readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations
that result from the gage or the connector face not being exactly perpendicular to the center axis.
Summary of Contents for 85051B
Page 4: ...iv 85051B ...
Page 7: ...85051B 1 1 1 General Information ...
Page 12: ...1 6 85051B General Information Clarifying the Terminology of a Connector Interface ...
Page 13: ...85051B 2 1 2 Specifications ...
Page 19: ...85051B 3 1 3 Use Maintenance and Care of the Devices ...
Page 37: ...85051B 4 1 4 Performance Verification ...
Page 40: ...4 4 85051B Performance Verification Recertification ...
Page 41: ...85051B 5 1 5 Troubleshooting ...
Page 43: ...85051B 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Page 46: ...5 6 85051B Troubleshooting Contacting Agilent ...
Page 47: ...85051B 6 1 6 Replaceable Parts ...
Page 50: ...6 4 85051B Replaceable Parts Introduction Figure 6 1 Replaceable Parts ...