85032F
1-5
General Information
Preventive Maintenance
port. This will be a male short from the calibration kit.
A connector gage is referred to in terms of the connector that it measures. For instance, a
male connector gage has a female connector on the gage so that it can measure male
devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of these are described in Chapter 3,“Use, Maintenance, and Care of the Devices.”
Failure to detect and remove dirt or metallic particles on a mating plane surface can
degrade repeatability and accuracy, and can damage any connector mated to it.
Improper connections, resulting from pin depth values being out of the
observed
limits (see
Table 2-3 on page 2-4), or from poor connection techniques, can also damage these devices.
Table 1-2 Clarifying the Sex of Connectors: Examples
Terminology
Meaning
Short
–f–
Female short (female center conductor)
Short (f)
Male short (male center conductor) to be connected to female port
Summary of Contents for 85032F
Page 4: ...iv 85032F ...
Page 7: ...1 1 1 General Information ...
Page 12: ...1 6 85032F General Information Preventive Maintenance ...
Page 13: ...2 1 2 Specifications ...
Page 18: ...2 6 85032F Specifications Electrical Specifications ...
Page 19: ...3 1 3 Use Maintenance and Care of the Devices ...
Page 33: ...4 1 4 Performance Verification ...
Page 36: ...4 4 85032F Performance Verification Recertification ...
Page 37: ...5 1 5 Troubleshooting ...
Page 41: ...6 1 6 Replaceable Parts ...
Page 45: ...A 1 A Standard Definitions ...