![Agilent Technologies 85024A User'S And Service Manual Download Page 16](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925016.webp)
2-4
Chapter 2
Accessories
Probe Adapter
Probe Adapter
Description
The probe adapter fits over the tip of the probe and converts the probe input to a 50
type-N male
connector. The adapter is only used when performance testing the probe.
Operating Characteristics
Inspection
Periodically inspect the threads of the adapter for signs of wear and damage. Inspect the barrel of the
probe receptacle, making sure it is clean and free of grit. Clean the adapter threads or receptacle with
clean compressed air.
Nominal Input Impedance
50
Frequency Range
Same as the active probe (300 kHz to 3 GHz)