Specifications
Specifications
Frequency Range
and Attenuation
Attenuation Accuracy
Maximum Insertion Loss
(0.4 + 0.9f) dB
where “f” is the frequency in GHz
Attenuation Repeatability
0.01 dB, dc to 18.0 GHz
0.05 dB, 18.0 to 26.5 GHz
RF Power Handling
Capability
1 W average, 100 W peak with maximum pulse width of 10 microseconds
Solenoid Drive
For Agilent 8495K only.
Switching speed
Maximum 20 milliseconds including settling time
Frequency Range
dc to 26.5 GHz
Attenuation
0 to 70 dB in 10 dB steps
Attenuation Accuracy ( dB, referenced from 0 dB)
Attenuation
Selection (dB)
dc–6.0 GHz
6.0–12.4 GHz
12.4–18.0 GHz
18.0–26.5 GHz
10
0.3
0.4
0.5
0.7
20
0.5
0.5
0.6
0.8
30
0.6
0.7
0.8
1.0
40
0.7
0.9
1.1
1.5
50
0.8
1.0
1.2
1.6
60
1.0
1.3
1.4
1.9
70
1.1
1.5
1.7
2.3
Solenoids Drive
Coil Voltage
Switching Current
8495K
20 to 30 Vdc
125 mA at 24V (approx.)