3-4
Calibration Overview
Factory Calibrations
Factory Calibrations
The following calibrations are performed at the factory:
Mainframe Calibration
O/E Factory Wavelength Calibration
Mainframe Calibration
Mainframe calibration affects both optical and electrical measurements. Main-
frame calibration improves timebase accuracy. All timebase measurements
such as rise time, fall time, eye width, jitter, and so forth are affected by the
timebase accuracy.
The calibration factors are stored in the nonvolatile RAM of the instrument.
There is a switch on the back panel of the instrument that allows the main-
frame calibration to be protected or unprotected. Next to the switch there is a
Table 3-1. Factory Calibration Summary
Calibration
What is calibrated
Measurements
Affected
Recommended
Interval
Softkey Path
Mainframe
Calibration
Accuracy and
continuity of the
timescale
Channels affected:
optical & electrical. All
time base
measurements such as
rise time, fall time, eye
width, and jitter.
Annually at Agilent
service center or if
operating temp has
changed and remains
5
°
C or more from
calibration
temperature. See
service manual.
Utility
Calibrate
Calibrate frame
O/E Factory
Wavelength
Calibration
The photodetector
responsivity
Channels affected:
optical. Amplitude
accuracy of all optical
channel
measurements. Optical
power meter accuracy.
Annual factory re-
calibration of standard
wavelengths.
Not user accessible.
a
a. Refer to “O/E User-Wavelength Calibration” on page 3-9.
Summary of Contents for 83482A
Page 1: ...Agilent 83482A Optical Electrical Plug In Module User s Guide ...
Page 6: ......
Page 8: ......
Page 16: ......
Page 27: ...2 11 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Page 28: ...2 12 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Page 66: ......
Page 84: ......
Page 88: ......