5-18
Servicing
Troubleshooting
To check the A4 Power Supply Board Assembly
1
Check the 6.3A fuse in the rear-panel’s line module FL1. Refer to “To check the
line-power fuse” on page 5-19.
2
Remove the power supply cable from J2 on the A5 Laser Driver Board
Assembly. The end of this cable can be probed to measure all of the dc voltages
supplied in the instrument. The following list shows each wire color, its color
code, and its purpose:
• Red (2) wire: +15V
• White/red (92) wire: +5V
• Violet (7) wire: –15V
• Black (0) wire: ground
• White/black (90) wire: ground
3
If no voltages are detected, check that the ac line input voltage at the cable that
connects to J1 on the A4 Power Supply Board Assembly. Measure the line
voltage across the white/gray/red wire and the gray wire. If the ac line voltage
is present, replace the A4 Power Supply Board Assembly.
Summary of Contents for 83437A
Page 1: ...Agilent 83438A Erbium ASE Source User s Guide ...
Page 5: ...v The Agilent 83438A At a Glance Rear view of instrument ...
Page 8: ......
Page 10: ......
Page 24: ...2 4 Making Measurements Performing Stimulus Response Measurements ...
Page 41: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information ...
Page 48: ......
Page 54: ......
Page 61: ...5 7 Servicing General Information ...
Page 63: ...5 9 Servicing General Information ...
Page 79: ...5 25 Servicing Adjustment Procedure ...
Page 85: ...5 31 Servicing Replacing Instrument Assemblies Location of resistors R2 R8 and R9 ...
Page 92: ...5 38 Servicing Replaceable Parts ...
Page 94: ...5 40 Servicing Replaceable Parts ...
Page 96: ...5 42 Servicing Replaceable Parts ...
Page 98: ...5 44 Servicing Replaceable Parts ...
Page 100: ...5 46 Servicing Replaceable Parts ...
Page 106: ......