1-11
Getting Started
Entering Calibration Data into an Agilent 8702A/B
Entering Calibration Data into an Agilent 8702A/B
Before using your source or receiver with an Agilent 8702A/B lightwave com-
ponent analyzer, you must load the calibration data as described in this sec-
tion. Two procedures are provided in this section. The first procedure is the
standard method. You will most likely use it. However, if you have an GPIB
external disk drive, you can use the second procedure. For more information
on calibration data,
refer to “Entering Calibration Data into an Agilent 8702D”
.
To install from the front panel
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word
RECEIVER
for
SOURCE
and the word
RCVR
for
SRC
.
1
On the Agilent 8702A/B, press:
PRESET
CAL
,
CAL KITS & STDS
,
SOURCE [COEFF]
,
SRC COEFF
,
ENTER SRC COEFF
2
Use the available softkeys to enter the coefficients that are listed on the label
that is attached to the source.
3
Press
LABEL STD
, and enter a label for the calibration data.
4
Press
STD DONE
and then
SAVE SRC COEFF
.