Entering Calibration Data into an Agilent 8702D
To copy the data from the disk
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word
RECEIVER
for
SOURCE
and the word
RCVR
for
SRC
.
1
Locate the DOS
-
formatted calibration disk provided with the source.
2
Place the disk in the Agilent 8702D’s front-panel disk drive.
3
Press
CAL
,
CAL KITS & STDS
,
SOURCE STANDARDS
,
LOAD SRC DISK MENU
. This will
display the disk directory screen.
4
Use the arrow keys or front panel knob to select a cal data file.
5
Load the cal data file into either of two internal registers by pressing the
corresponding
LOAD
softkey.
To manually enter data from the label
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word
RECEIVER
for
SOURCE
and the word
RCVR
for
SRC
.
1
Press
CAL
,
CAL KITS & STDS
,
SOURCE
,
CAL STD:SRC COEFF
and then
ENTER SRC COEFF
.
2
Press each softkey corresponding to the nine coefficients (A through I), using
the
MORE
softkey, and enter the coefficients. The coefficients are listed on a
label that is on the source. After all of the coefficients are entered, press
RETURN
.
3
Press
LABEL STD
.
4
Use the front-panel knob and the
SELECT LETTER
softkey to write a title for the
calibration data.
5
Press
STD DONE (DEFINED)
and then
SAVE SRC COEFF
.
6
Press
CAL
,
CALIBRATE MENU
,
RESPONSE
, and then
SOURCE
.
7
When the
SOURCE
softkey is underlined, press
DONE: RESPONSE
.
Summary of Contents for 83400 Series
Page 2: ...Agilent 83400 Series Lightwave Source and Receiver Modules User s Guide ...
Page 6: ...v Laser Classification IEC and FDA labels and their placement on an Agilent 83402C ...
Page 9: ......
Page 11: ......
Page 17: ...1 6 Getting Started Connecting a Source Figure 1 2 Location of laser safety labels ...
Page 81: ...Index 4 Index width spectral procedure 4 4 Y YAG system dual hetrodyne 4 7 ...