Chapter 4, GPIB Commands
TX Subsystem
TX:INPut:POWer:TRIGger[:SOURce]
ëû
{AMPTD|EXT|IMM}
Selects the trigger source to initiate the power measurement after the
TX:INP:POW?
command is sent.
AMPTD:
Executes the power measurement when a signal greater than
−
5 dBm
is input to the RF IN/OUT port (amplitude trigger).
EXT:
Executes the power measurement when a TTL level signal is input to
the EXT TRIG IN port (external trigger).
IMM:
Executes the power measurement immediately (immediate trigger).
•
Response to the Query
{AMPTD | EXT | IMM} <new line><^END>
TX:INPut:POWer:UNIT
ëû
{DBM|W}
Specifies the measurement units for the input signal’s power result returned by
the
TX:INP:POW?
command.
DBM:
dBm
W:
W (Watt)
•
Response to the Query
{DBM | W} <new line><^END>
TX:OUTPut[:RFANalyzer]:ATTenuator
ëû
<numeric>
[DB]
Specifies the TX path attenuator value.
<numeric>:
0 to 40 dB in 1 dB steps.
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
NOTE:
This command is not generally needed to set the level because the attenuator is
automatically set when using the
TX:OUTP:LEV:ADJ
command.
•
Response to the Query
{numeric (dB)} <new line><^END>
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...