
117
Chapter 4, GPIB Commands
TX Subsystem
TX:INPut:POWer:PDETector:SAMPle:IGNore
ëû
<
numeric
>
Specifies the number of the A/D conversion samples to ignore after the positive
edge trigger. (The sampling time is 12.5
µ
s.)
This command is valid only when the
TX:INP:POW:TRIG
command is set to
AMPTD
or
EXT
.
<numeric>:
3 to 3200.
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric} <new line><^END>
NOTE:
The following equation shows the relationship between the IGNORE value set by the
TX:INP:POW:PDET:SAMP:IGN
command and the LENGTH value set by the
TX:INP:POW:PDET:SAMP:LENG
command.
IGNORE value + LENGTH value
≤
4800
TX:INPut:POWer:PDETector:SAMPle:LENGth
ëû
<
numeric
>
Specifies the number of A/D conversion samples to be used to calculate the
measurement value of the power measurement. (The sampling time is 12.5
µ
s.)
<numeric>:
1 to 4800
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric} <new line><^END>
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...