103
Chapter 4, GPIB Commands
COMPensation Subsystem
COMPensation[:RFG1]:TEMPerature:LEVel?
This command sets the signal path for the temperature compensation, and returns
the level required at the FROM DUPLEX OUT port for the temperature
compensation for the RF signal generator to provide optimum output level
accuracy. The band must be specified by executing
the
COMP:TEMP:FREQ?
ëû
{0|1|2|3|4|5|<numeric>}
to specify the frequency before executing this command
.
(Query only)
Supply the signal whose frequency and level are returned by this command and
the
COMP:TEMP:FREQ?
ëû
{0|1|2|3|4|5|<numeric>}
command to the FROM
DUPLEX OUT port. Then execute the
COMP:TEMP:EXEC?
command to
execute the RF generator temperature compensation.
•
Response to the Query
{numeric (dBm)} <new line><^END>
COMPensation[:RFG1]:TEMPerature:PDETector:SAMPle:AVERage[:FACTor]
ëû
<
numeric
>
Specifies the number of bursts to be measured to calculate the average value of
the dc offset measurement.
<numeric>:
1 to 10
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric} <new line><^END>
COMPensation[:RFG1]:TEMPerature:PDETector:SAMPle:LENGth
ëû
<
numeric
>
Specifies the number of A/D conversion samples to be used to calculate the
measurement value of the dc offset measurement. (The sampling time is 12.5
µ
s.)
<numeric>:
1 to 4800
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric} <new line><^END>
Summary of Contents for 83236B
Page 7: ...vii...
Page 10: ...x...
Page 14: ...Contents 14 Error Message 154 Glossary 161...
Page 30: ...30 Chapter 1 General Information Other Information...
Page 123: ...123 5 S pecifications Specifications...
Page 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Page 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Page 141: ...141 B M anual Changes Appendix B Manual Changes...