Operation
4
SCD and NCD User Manual
57
Configure Auto Flow Zero on the GC
Agilent recommends setting the GC to automatically zero flow
sensors to reduce drift. See the GC
Operation Manual
for
details.
1
On the GC keypad, press
[Options]
.
2
Scroll to
Calibration
and press
[Enter]
.
3
Scroll to select the appropriate detector (front, back, aux 2,
or aux 1) and press
[Enter].
4
Scroll to
Autoflow zero (H2 Lower)
and press [
On/Yes
]. (To turn
off autozero, instead press [
Off/No
].
5
For SCD, repeat for
Autoflow zero (H2 Upper)
.